FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 2872108 · Received December 6, 2012

Report

Report Number
1222780-2012-00262
Event Type
Injury
Date Received
December 6, 2012
Date of Event
November 9, 2012
Report Date
November 9, 2012
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SERIAL NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT. SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. THE DEVICE HAS NOT YET BEEN RETURNED THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. IT THE DEVICE IS RETURNED AND EVAL COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGEMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. REF INTERNAL COMPLAINT (B)(6).

Description of Event or Problem · 1

FOLLOWING SEVERAL UNSUCCESSFUL CAVITY INTEGRITY ASSESSMENT (CIA) TESTS WITH 3 DISPOSABLE NOVASURE DEVICES THE PHYSICIAN ABORTED THE PROCEDURE AND DID A LAPAROSCOPIC TUBAL LIGATION. A UTERINE PERFORATION WAS SEEN ¿IN THE FUNDUS¿ AT THAT TIME. NO TREATMENT WAS NEEDED AND THE PT WAS DISCHARGED HOME. A HYSTERECTOMY, DILATATION AND CURETTAGE (D&C), AND SOUNDING (NOT HOLOGIC DEVICES) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 12G13R

Patients

Seq Age Sex Outcome Treatment
1 UNK Other RADIO FREQUENCY CONTROLLER - SERIAL # UNK