FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, SINGLE CONN./EXT. DL

MDR report key: 2872063 · Received December 11, 2012

Report

Report Number
8030665-2012-00381
Event Type
Malfunction
Date Received
December 11, 2012
Date of Event
November 13, 2012
Report Date
November 13, 2012
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE COMPLAINT IS CONFIRMED. AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCE DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

RN HAS REPORTED THAT DIALYSIS SOLUTION IS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLE. THE LEAK WAS NOTED AT THE POINT WHERE THE TUBING CONNECTS WITH THE CASSETTE. THERE IS NO REPORT OF PATIENT ILL EFFECT. SAMPLE IS AVAILABLE FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIBERTY CYCLER SET, SINGLE CONN./EXT. DL LIBERTY DIALYSIS CYCLER TUBING FKX REYNOSA MANUFACTURING 11SR08043

Patients

Seq Age Sex Outcome Treatment
1 LIBERTY DIALYSIS CYCLER