PERMANENT CAUTERY SPATULA INSTRUMENT
Report
- Report Number
- 2955842-2012-01240
- Event Type
- Malfunction
- Date Received
- December 11, 2012
- Date of Event
- November 12, 2012
- Report Date
- November 12, 2012
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE CERAMIC SLEEVE HAS A SECTION BROKEN OFF AT THE DISTAL END. PIECES OF CERAMIC ARE MISSING FROM THE SPATULA, EXPOSING THE SPATULA SHANK. THE SLEEVE ALSO EXHIBITS SCRATCH MARKS BELOW THE BROKEN SECTION. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. INTUITIVE SURGICAL CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS OF THE DATE OF THIS REPORT.
IT WAS REPORTED THAT DURING REPROCESSING OF THE PERMANENT CAUTERY SPATULA INSTRUMENT, THE INSULATION ON THE INSTRUMENT WAS CRACKED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERMANENT CAUTERY SPATULA INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL, INC. | 420184-06 | M10110304 123 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| DA VINCI SI SYSTEM INSTRUMENTS |