FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY SPATULA INSTRUMENT

MDR report key: 2871998 · Received December 11, 2012

Report

Report Number
2955842-2012-01240
Event Type
Malfunction
Date Received
December 11, 2012
Date of Event
November 12, 2012
Report Date
November 12, 2012
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE CERAMIC SLEEVE HAS A SECTION BROKEN OFF AT THE DISTAL END. PIECES OF CERAMIC ARE MISSING FROM THE SPATULA, EXPOSING THE SPATULA SHANK. THE SLEEVE ALSO EXHIBITS SCRATCH MARKS BELOW THE BROKEN SECTION. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. INTUITIVE SURGICAL CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS OF THE DATE OF THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE PERMANENT CAUTERY SPATULA INSTRUMENT, THE INSULATION ON THE INSTRUMENT WAS CRACKED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERMANENT CAUTERY SPATULA INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL, INC. 420184-06 M10110304 123

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| DA VINCI SI SYSTEM INSTRUMENTS