PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2012-03330
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 20, 2012
- Report Date
- November 20, 2012
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT ANALYSIS WAS APPROVED ON (B)(6) 2013. AN ANALYSIS WAS PERFORMED ON THE RETURNED FLASHCARD, AND NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. AN ANALYSIS WAS PERFORMED ON THE RETURNED HANDHELD AND THE REPORTED ALLEGATION OF "FAILURE TO POWER ON OR OFF" WAS NOT VERIFIED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD PERFORMANCE WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.
A VNS PROGRAMMING PHYSICIAN CONTACTED OUR CONSULTANT IN (B)(6). IT WAS REPORTED THAT THEIR HANDHELD COMPUTER DOESN'T SWITCH ON ANYMORE. THE HANDHELD WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | 250 | 874820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |