FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2871856
·
Received December 13, 2012
Report
- Report Number
- 3004209178-2012-91337
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- October 31, 2012
- Report Date
- November 30, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH AN ALARM AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED CONSTANTLY. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722CAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |