FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2871840 · Received December 13, 2012

Report

Report Number
3004209178-2012-91317
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
September 30, 2012
Report Date
November 30, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED. ADVISED THE CALLER THAT THE DEVICE WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 52 YR