FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2871838 · Received December 13, 2012

Report

Report Number
3004209178-2012-91315
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
September 26, 2012
Report Date
December 5, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. HOWEVER, THE DEVICE ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE INSULIN PUMP HAD A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, BROKEN BATTERY TUBE THREADS, AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR SEVERAL TIMES. IT WAS STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD, BUT IT WAS CONTACT WITH X-RAYS. THE DISPLACEMENT TEST WAS PERFORMED AND PASSED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWP

Patients

Seq Age Sex Outcome Treatment
1