ANIMAS INSULIN INFUSION PUMP
MDR report key: 2871751
·
Received December 13, 2012
Report
- Report Number
- 2531779-2012-14528
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- September 27, 2012
- Report Date
- October 17, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. ALL OF THE KEYPAD BUTTONS WERE FOUND TO BE RESPONDING PROPERLY TO BUTTON PRESSES. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE ISSUE, THE VIBRATION MOTOR WAS FOUND TO HAVE FAILED. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR A DAMAGED KEYPAD. EVALUATION FOUND NO DAMAGE TO THE KEYPAD BUT FOUND CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION.