FDA Adverse Event Malfunction Summary report: N

ELECSYS 2010 DISK

MDR report key: 2871632 · Received December 13, 2012

Report

Report Number
1823260-2012-06315
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 27, 2012
Report Date
January 16, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K961481
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A ROOT CAUSE COULD NOT BE DETERMINED. CALIBRATION AND QUALITY CONTROL WERE GOOD PRIOR TO THE RUN. A PERFORMANCE TEST WAS DONE AND SHOWED NO INDICATION OF AN INSTRUMENT ISSUE. A SERVICE VISIT WAS PERFORMED. A COMPLETE MAINTENANCE WAS DONE AND THE SAMPLE PIPETTOR WAS CHANGED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURED IN (B)(6).

Description of Event or Problem · 1

THE CUSTOMER ALLEGED THEY RECEIVED QUESTIONABLE N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE (PROBNP) RESULTS ON THEIR ELECSYS 2010 ANALYZER FOR TWO PATIENTS. THE INITIAL PROBNP RESULTS FOR BOTH PATIENTS WERE NOT CONSISTENT WITH THE PATIENTS' CONDITIONS AND NEW MEASUREMENTS WERE REQUESTED. THE FIRST PATIENT'S INITIAL PROBNP RESULT WAS 46.74 PG/ML AND IT WAS REPORTED OUTSIDE THE LABORATORY. ON (B)(6) 2012, THE SAMPLE WAS REPEATED AND THE RESULT WAS 10207 PG/ML. ON (B)(6) 2012, THE PATIENT HAD A NEW SAMPLE DRAWN AND THE RESULT WAS 7509 PG/ML. ON (B)(6) 2012, THE SECOND PATIENT, AN (B)(6), MALE, HAD AN INITIAL PROBNP RESULT OF 49.51 PG/ML AND IT WAS REPORTED OUTSIDE THE LABORATORY. ON (B)(6) 2012, THE SAMPLE WAS REPEATED AND THE RESULT WAS 21065 PG/ML. ON (B)(6) 2012, THE PATIENT HAD ANOTHER SAMPLE TUBE TESTED THAT WAS DRAWN AT THE SAME TIME AS THE ORIGINAL SAMPLE AND THE RESULT WAS 21631 PG/ML. THE PATIENTS WERE NOT ADVERSELY AFFECTED BY THIS EVENT. THE PROBNP REAGENT LOT NUMBER WAS 168452 AND THE EXPIRATION DATE WAS NOT PROVIDED.