FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2871535 · Received December 13, 2012

Report

Report Number
2531779-2012-14519
Event Type
Malfunction
Date Received
December 13, 2012
Report Date
November 15, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4): TESTING CONFIRMED INTERMITTENT RESPONSE FROM THE OK AND UP BUTTONS. THE KEYPAD WAS PEELING AND WAS REMOVED: FOUND CONTAMINATION UNDER ALL CONTACTS. UNRELATED TO THIS COMPLAINT, THE DISPLAY IS DIM/FADING. WHEN REPLACED WITH A TEST SCREEN , IT FUNCTIONED PROPERLY.

Description of Event or Problem · 1

ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD ISSUE OF MULTIPLE UNRESPONSIVE KEYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1