ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-14519
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Report Date
- November 15, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).
(B)(4): TESTING CONFIRMED INTERMITTENT RESPONSE FROM THE OK AND UP BUTTONS. THE KEYPAD WAS PEELING AND WAS REMOVED: FOUND CONTAMINATION UNDER ALL CONTACTS. UNRELATED TO THIS COMPLAINT, THE DISPLAY IS DIM/FADING. WHEN REPLACED WITH A TEST SCREEN , IT FUNCTIONED PROPERLY.
ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD ISSUE OF MULTIPLE UNRESPONSIVE KEYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.