FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2871450 · Received December 13, 2012

Report

Report Number
1416980-2012-07191
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 28, 2012
Report Date
November 28, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS NOT AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE PROBLEM WAS CONFIRMED FOR USE ERROR-FAILED TO FOLLOW THERAPY STEPS ON THE PHONE. THE ASSIGNABLE CAUSE OF THE PROBLEM IS THE PATIENT INAPPROPRIATELY BYPASSED THE DRAIN PHASES WHICH IS A USE ERROR. REVIEW OF THE SERVICE HISTORY REVEALED NO FAILURES/PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY AND THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. REVIEW OF THE DEVICE SERVICE HISTORY REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY OF USE ERROR-FAILED TO FOLLOW THERAPY STEPS.

Description of Event or Problem · 1

THE HOME PATIENT (HP) CONTACTED BAXTER'S SERVICE CENTER REPORTING AN ALARM THAT OCCURRED ON THE HOMECHOICE (HC) DURING USE IN DRAIN 4 OF 4 AND THE HP STATED SHE BYPASSED EVERY DRAIN. THE TSR HAD THE HP CHECK FOR KINKS, CLOSED CLAMPS AND FIBRIN. THERE WERE NO PROBLEMS FOUND. THE HP REPOSITIONED AND THE ALARM REOCCURRED. THE TSR REVIEWED THE PROGRAM FOR CONTINUOUS CYCLING PERITONEAL DIALYSIS (CCPD). THE HP REQUESTED TO BYPASS AND THE TSR ASSISTED AS REQUESTED. THE TSR ASSISTED TO END THERAPY. THE HP HAD AN APPOINTMENT WITH THE REGISTERED NURSE (RN) AND TSR ADVISED THE HP TO LET THE RN KNOW OF THE ALARM. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 63 YR CASSETTE