HOMECHOICE PRO
Report
- Report Number
- 1416980-2012-07191
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 28, 2012
- Report Date
- November 28, 2012
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE DEVICE IS NOT AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SUBMITTED.
(B)(4). THE PROBLEM WAS CONFIRMED FOR USE ERROR-FAILED TO FOLLOW THERAPY STEPS ON THE PHONE. THE ASSIGNABLE CAUSE OF THE PROBLEM IS THE PATIENT INAPPROPRIATELY BYPASSED THE DRAIN PHASES WHICH IS A USE ERROR. REVIEW OF THE SERVICE HISTORY REVEALED NO FAILURES/PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY AND THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. REVIEW OF THE DEVICE SERVICE HISTORY REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY OF USE ERROR-FAILED TO FOLLOW THERAPY STEPS.
THE HOME PATIENT (HP) CONTACTED BAXTER'S SERVICE CENTER REPORTING AN ALARM THAT OCCURRED ON THE HOMECHOICE (HC) DURING USE IN DRAIN 4 OF 4 AND THE HP STATED SHE BYPASSED EVERY DRAIN. THE TSR HAD THE HP CHECK FOR KINKS, CLOSED CLAMPS AND FIBRIN. THERE WERE NO PROBLEMS FOUND. THE HP REPOSITIONED AND THE ALARM REOCCURRED. THE TSR REVIEWED THE PROGRAM FOR CONTINUOUS CYCLING PERITONEAL DIALYSIS (CCPD). THE HP REQUESTED TO BYPASS AND THE TSR ASSISTED AS REQUESTED. THE TSR ASSISTED TO END THERAPY. THE HP HAD AN APPOINTMENT WITH THE REGISTERED NURSE (RN) AND TSR ADVISED THE HP TO LET THE RN KNOW OF THE ALARM. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO SERIOUS INJURY OR MEDICAL INTERVENTION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | CASSETTE |