FDA Adverse Event Malfunction Summary report: N

ULTRACISION** HARMONIC SCALPEL** CURVED SHEARS, SCISSOR-GRIP HANDLE - 14 CM

MDR report key: 2871449 · Received December 13, 2012

Report

Report Number
3005075853-2012-05634
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 27, 2012
Report Date
November 30, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH TISSUE PAD DETACHED BUT WITH EVIDENCE OF THE TISSUE PAD MATERIAL IN THE GROOVE SECTION OF THE CLAMP ARM THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE FOR THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. BOTH CONDITIONS MAY CAUSE A SYSTEM FAILURE SIGNALED BY A CONTINUOUS BEEP WHEN EITHER OF THE FOOT PEDALS IS DEPRESSED. KEEP THE CLAMP ARM OPEN WHEN BACKCUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU, CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE.

Additional Manufacturer Narrative · 1

(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CORONARY ARTERY BYPASS GRAFT, THE TISSUE PAD WAS DETACHED OFF. THE TISSUE PAD WAS RETRIEVED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION** HARMONIC SCALPEL** CURVED SHEARS, SCISSOR-GRIP HANDLE - 14 CM ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK J4AR3J

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE