FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2871169 · Received December 12, 2012

Report

Report Number
3008382007-2012-07940
Event Type
Malfunction
Date Received
December 12, 2012
Report Date
December 4, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (08/12/2013)-DEVICE EVALUATION: THE ANALYSIS OF THE SUBJECT METER WAS COMPLETED ON (B)(6) 2013 BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE REPORTED COMPLAINT COULD NOT BE REPRODUCED DURING INVESTIGATION. THE METER FUNCTIONED NORMALLY AND NO ISSUES WERE FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIO IQ METER READ INACCURATELY COMPARED TO A LABORATORY DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "297 MG/DL" WITH THE SUBJECT METER AND "242 MG/DL" ON A LABORATORY DEVICE, PERFORMED WITHIN 10-30 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO LAB ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3376683

Patients

Seq Age Sex Outcome Treatment
1 79 YR