FDA Adverse Event Malfunction Summary report: N

SUCTION / IRRIGATOR INSTRUMENT

MDR report key: 2871116 · Received December 12, 2012

Report

Report Number
2955842-2012-01286
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
August 28, 2012
Report Date
September 13, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K110451
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT OF THE TIP ALMOST CAME OFF IS CONFIRMED. INSTRUMENT WAS RETURNED WITH ALL THE DISTAL END CABLES CUT AND THE SNAKE WRIST AND TIP ASSEMBLY DETACHED FROM THE MAIN TUBE. SNAKE WRIST DISKS ARE NOT IN ALIGNMENT. EVIDENCE NOT CONCLUSIVE, BUT LIKELY FAILURE WAS A DISLODGED SNAKE WRIST DISK. WRIST CABLES WERE LIKELY CUT TO ENABLE REMOVAL OF INSTRUMENT FROM CANNULA. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI LOBECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED THAT THE TIP OF THE SUCTION IRRIGATOR INSTRUMENT ALMOST CAME OFF. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUCTION / IRRIGATOR INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 410299-02 M10120510 183

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES