HARMONIC ACE CURVED SHEARS INSERT
Report
- Report Number
- 2955842-2012-01258
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 16, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K093217
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE INSERT ACCESSORY AND HARMONIC ACE INSTRUMENT WERE RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSERT ACCESSORY WAS RETURNED INSTALLED ON THE HARMONIC ACE INSTRUMENT. THE INSERT ACCESSORY IS NOT FULLY INSTALLED INTO INSTRUMENT, AS THE INSTRUMENT GEARS ARE NOT MESHING WITH THE INSERT ACCESSORY'S ROUND RACK FEATURE. DISASSEMBLY OF THE INSTRUMENT FOUND THAT THE INSERT IS INCORRECTLY ORIENTED WITHIN THE MAIN TUBE AND THE PEAK ON INSERT ADAPTER IS MAKING CONTACT WITH PEAK ON THE MATING MAIN TUBE FEATURE, THUS PREVENTING PROPER INSTALLATION OF INSERT. THE INSERT PEAK FEATURE SHOULD BE ORIENTED 180 DEGREES FROM THE MAIN TUBE PEAK FEATURE. ENGINEERING EVALUATION ALSO FOUND THAT THE DISTAL END OF THE INSERT IS MISSING THE CLAMP ARM. ONE SIDE OF THE INSERT THAT HOLDS CLAMP ARM PIN IS BENT AND THE HOLE FOR THE PIN IS BROKEN AT ONE EDGE. ENGINEERING CONCLUDED THAT DAMAGE TO THE CLAMP ARM IS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. ON (B)(4) 2012, INTUITIVE SURGICAL CONTACTED (B)(6) (INITIAL REPORTER), THE MATERIALS COORDINATOR AT (B)(6) HOSPITAL IN (B)(6). TINA INDICATED THAT IT WAS NOTED THAT DURING DISASSEMBLY OF THE HARMONIC ACE INSTRUMENT AND INSERT ACCESSORY, THE INSERT ACCESSORY WAS STUCK AND COULD NOT BE REMOVED. (B)(6) INDICATED THAT THERE WAS NO REPORT BY THE SURGICAL STAFF THAT THE DEVICE FAILED DURING THE SURGICAL PROCEDURE AND THAT NO PIECE FROM THE INSERT ACCESSORY WAS REPORTED TO HAVE FALLEN INTO THE PATIENT.
IT WAS REPORTED THAT AFTER COMPLETION OF A DA VINCI SI SURGICAL PROCEDURE, THE SITE WAS UNABLE TO REMOVE THE HARMONIC ACE CURVED SHEARS INSERT ACCESSORY INSTALLED ON THE HARMONIC ACE INSTRUMENT. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE CURVED SHEARS INSERT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400272-02 | M10120213 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM, ACCESSORIES AND ESU. |