ILAB ULTRASOUND IMAGING SYSTEM
Report
- Report Number
- 2134265-2012-08210
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Date of Event
- November 14, 2012
- Report Date
- November 14, 2012
- Manufacturer
- BOSTON SCIENTIFIC - FREMONT (CE)
- Product Code
- OBJ
- PMA / PMN Number
- K051679
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATED BY MFR: EXAMINATION OF THE RETURNED DEVICE OBSERVED WITH NO VISIBLE DAMAGE OR DEFECTS. THE DEVICE MEETS SPECIFICATIONS FOR FUNCTIONAL TEST AND SUCCESSFULLY UNDERWENT A CONTINUOUS BURN-IN FOR 6 HOURS. THE REPORTED PROBLEM COULD NOT BE REPRODUCED AS INDICATED IN THE COMPLAINT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS NOT CONFIRMED. (B)(4).
(B)(4). DEVICE EVALUATED BY MFR: IT IS INDICATED THAT THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
IT WAS REPORTED THAT DURING AN INTRAVASCULAR ULTRASOUND DIAGNOSIS PROCEDURE, THE ILAB ULTRASOUND IMAGING SYSTEM MOTORDRIVE AUTOMATIC PULLBACK FUNCTION STOPPED WORKING DURING THE PROCEDURE. THE MANUAL PULLBACK OF THE IMAGING CATHETER WAS PERFORMED WHICH DID NOT GIVE GOOD RESULT. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS GOOD.
IT WAS REPORTED THAT DURING AN INTRAVASCULAR ULTRASOUND DIAGNOSIS PROCEDURE, THE ILAB ULTRASOUND IMAGING SYSTEM MOTORDRIVE AUTOMATIC PULLBACK FUNCTION, STOPPED WORKING DURING THE PROCEDURE. THE MANUAL PULLBACK OF THE IMAGING CATHETER WAS PERFORMED WHICH DID NOT GIVE GOOD RESULT. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ILAB ULTRASOUND IMAGING SYSTEM | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | BOSTON SCIENTIFIC - FREMONT (CE) | H749ILAB240INS0 | 0000006171 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |