FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2870779 · Received December 12, 2012

Report

Report Number
2531779-2012-14490
Event Type
Malfunction
Date Received
December 12, 2012
Report Date
November 14, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2012, WITH THE FOLLOWING FINDINGS: TESTING WAS UNABLE TO DUPLICATE LARGE PRIME VOLUME. MULTIPLE OCCURRENCES OF HIGH PRIME VOLUME OBSERVED IN HISTORY. FORCE CALIBRATION FAILED WITH A LOW READING. .FORCE SENSOR RESISTANCE TEST FAILED WITH A HIGH READING. DISASSEMBLED FORCE SENSOR, NO EVIDENCE OF SHIM OR PLATE DAMAGE. EVIDENCE OF CONTAMINATION FOUND IN FORCE SENSOR HOUSING AND FORCE SENSOR SHIM. REMOVED PUMP CASE AND NOTED FORCE SENSOR PINS SEATED AND SOLDER CONNECTIONS INTACT. NO EVIDENCE OF MOISTURE CONTAMINATION ON PINS OR IN PUMP.

Description of Event or Problem · 1

THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED CONTAMINATION ON FORCE SENSOR, FORCE SENSOR CALIBRATION LOW AND FORCE SENSOR RESISTANCE HIGH. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 48 YR