FDA Adverse Event Injury Summary report: N

ARTICUL/EZE FEMORAL HEADS

MDR report key: 2870562 · Received December 12, 2012

Report

Report Number
1818910-2012-83564
Event Type
Injury
Date Received
December 12, 2012
Date of Event
December 5, 2012
Report Date
July 24, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
PK910583
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

**UPDATE: (B)(4) 2013 - LITIGATION PAPERS RECEIVED. LITIGATION ALLEGES DISCOMFORT. DEPUY CONSIDERS THIS COMPLAINT CLOSED AT THIS TIME.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. C. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS HIGH CHROMIUM ION LEVEL AND PAIN. INTRAOPERATIVELY, THE FEMORAL HEAD WAS FOUND TO BE SCRATCHED, AND THE SUPERIOR EDGE OF THE LINER WAS WORN, BELIEVED TO BE CAUSED BY SUBLUXATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARTICUL/EZE FEMORAL HEADS FEMORAL HEAD JDI DEPUY ORTHOPAEDICS INC US 2952977

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention