UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2012-02823
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 14, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ADDITIONAL INFORMATION: THE FRONT OF THE WHITE BLOOD CELL (WBC) HISTOGRAM SHOWS A MODERATE INTERFERENCE PATTERN. THE NUCLEATED RED BLOOD CELL (NRBC) SCATTER PLOT HAS SOME DEBRIS EVENTS, BUT NOT ENOUGH TO TRIGGER THE PLATELET CLUMP FLAG. AS A RESULT, THE ALGORITHM DID NOT FLAG FOR CLUMPS. THERE WAS NOT SIGNIFICANT INTERFERENCE OF DEBRIS NOTED ON WBC HISTOGRAM OR NRBC MODULE (FOR PLATELET EVENTS). THE CUSTOMER ALSO PERFORMED A MANUAL WBC CORRECTION USING THE OBSERVED PLATELET (PLT) CLUMPS ON THE SLIDE. A REVIEW OF THE WBC HISTOGRAM DID NOT INDICATE A SIGNIFICANT AMOUNT OF THE DEBRIS. IN ADDITION, THE NRBC MODULE HAS ONLY APPROXIMATELY 4% OF PLATELET CLUMPS. THEREFORE, THE MANUAL CORRECTION ON THE WBC MAY NOT HAVE BEEN NEEDED.
THE ACTUAL DEVICE WAS NOT EVALUATED. THE CUSTOMER FORWARDED A PRINTOUT OF THE RESULTS FROM THE SYSTEM FOR THIS EVENT. THE PRINTOUT WAS REVIEWED; HOWEVER, NO CONCLUSION CAN BE DRAWN FROM THE INFORMATION PROVIDED. PER LABELING: GIANT PLATELETS, PLATELET CLUMPS, WHITE CELL FRAGMENTS, ELECTRONIC NOISE, VERY SMALL RED CELLS, AND RED CELL FRAGMENTS ARE LISTED AS INTERFERING SUBSTANCES FOR PLATELET PARAMETER.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT ERRONEOUS LOW PLATELET (PLT) RESULTS WERE GENERATED BY THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM WITHOUT INSTRUMENT GENERATED FLAGS, FOR A PATIENT SPECIMEN CONTAINING PLATELET CLUMPS. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LAB. THE LOW PLATELET VALUE PROMPTED THE CUSTOMER TO PERFORM A MANUAL REVIEW WHERE THE PLATELET CLUMPS WERE DETECTED (115 PLT CLUMPS/100 WHITE BLOOD CELLS (WBC). THE WBC RESULT WAS THEN CORRECTED FROM 14.5 TO 6.7. ACCORDINGLY, THE WBC RESULT WAS CONSIDERED ERRONEOUSLY HIGH, WITH NO INSTRUMENT FLAGS. ADDITIONALLY, AN ACCURATE MANUAL PLATELET COUNT COULD NOT BE OBTAINED DUE TO THE PLATELET CLUMPING. THERE ARE NO REPORTS OF ADVERSE EVENTS OR THE NEED FOR ANY MEDICAL INTERVENTION TO PREVENT AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |