FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2870299 · Received December 12, 2012

Report

Report Number
2531779-2012-14472
Event Type
Malfunction
Date Received
December 12, 2012
Report Date
November 15, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON 11/15/2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. DURING EVALUATION THE PUMP WAS ABLE TO COMPLETE THE REWIND, LOAD AND PRIME STEPS SUCCESSFULLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A - 2531779-03/24/2010-003-R.

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012, WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THIS REPORT IS BEING MADE BASED ON EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 39 YR