PULSAR II GENERATOR
Report
- Report Number
- 1226420-2012-00098
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Date of Event
- November 13, 2012
- Report Date
- March 18, 2013
- Manufacturer
- MEDTRONIC ADVANCED ENERGY, LLC
- Product Code
- MUL
- PMA / PMN Number
- K102029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION PROCESS: UNIT RECEIVED IN GOOD CONDITION AND INTERNAL VISUAL INSPECTION REVEALED NO LOOSE OR BROKEN COMPONENTS. DISTORTED WAVEFORMS EVIDENT IN BOTH CUT AND COAG MODES. FOUND R7 (5.1OHM) READING OPEN CIRCUIT AND R8 (5.1OHM) READING 1.2KOHMS ON RF BOARD. AFTER RESOLDERING R7 AND R8 ON THE RF BOARD THE UNIT DELIVERED RF ENERGY CORRECTLY INTO FIXED RESISTORS. RF AMPLIFIER PCBA IS BUILT ON OBSOLETE PCB AND DC POWER SUPPLY PCBA IS BUILT ON OBSOLETE PCB AND DUE TO THE COST OF UPGRADING THIS UNIT THE UNIT WILL BE SCRAPPED. ROOT CAUSE: COMPONENTS R7 AND R8 ON THE RF AMP BOARD HAD CRACKED/COLD SOLDER JOINTS PREVENTING THE OUTPUT TRANSFORMER FROM BEING DRIVEN PROPERLY. (B)(4).
(B)(4). EVAL METHOD, RESULTS AND CONCLUSION: PRODUCT SCHEDULED FOR RETURN BUT NOT RECEIVED BY MANUFACTURER FOR INSPECTION. PRODUCT EVENT: (B)(6).
A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
GENERATOR WORKING SPORADICALLY, FAILURE TO CUT OR COAG, NO POWER TO DEVICE. ELECTROCAUTERY UTILIZED TO COMPLETE CASE.
GENERATOR WORKING SPORADICALLY, FAILURE TO CUT OR COAG, NO POWER TO DEVICE. ELECTROCAUTERY UTILIZED TO COMPLETE CASE.