FDA Adverse Event Malfunction Summary report: N

PROGRASP (TM) FORCEPS

MDR report key: 2870244 · Received December 12, 2012

Report

Report Number
2955842-2012-01254
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
October 24, 2012
Report Date
November 15, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMPS WAS MISSING IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS MAIN INSULATION TUBE DAMAGE. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS AND DAMAGE NEAR THE DISTAL END. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE 'PROGRASP FORCEPS INSTRUMENT WOULD NOT CLOSE(NOTHING FELL INTO THE PATIENT, NO PATIENT HARM).'

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRASP (TM) FORCEPS ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M10120906 379

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES