FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2870058 · Received December 12, 2012

Report

Report Number
2032227-2012-07996
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
November 1, 2012
Report Date
November 1, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

UP ARROW BUTTON UNRESPONSIVE DUE TO CORRODED KEYPAD TRACES. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND NO DELIVERY.

Description of Event or Problem · 1

THE CUSTOMER CALLED BECAUSE, SHE FELT THAT THE NO DELIVERY ALARM WAS NOT WORKING. ADVISED THE CUSTOMER THAT TROUBLESHOOTING COULD BE CONDUCTED TO TEST THE ALARM'S FUNCTION, BUT SHE DIDN'T HAVE THE TUBING CLAMP. THE TUBING CLAMP WAS SHIPPED. THE CUSTOMER LATER CALLED TO REPORT UNRESPONSIVE BUTTONS. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-522LNAP

Patients

Seq Age Sex Outcome Treatment
1 25 YR