FDA Adverse Event Malfunction Summary report: N

PROGRASP (TM) FORCEPS

MDR report key: 2870036 · Received December 12, 2012

Report

Report Number
2955842-2012-01248
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
August 30, 2012
Report Date
September 14, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING FOUND THE DAMAGED INSTRUMENT COMPONENT WAS A BROKEN PITCH CABLE. THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN CLEVIS. CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING AND STERILIZATION IN CENTRAL PROCESSING IT WAS NOTED THAT THE PROGRASP (TM) FORCEPS HAD A LOOSE CABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRASP (TM) FORCEPS ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M20120319 799

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES