ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14464
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Report Date
- November 14, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP #1 DATE OF SUBMISSION 01/29/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: ACCORDING THE BLACK BOX, ONE NO CARTRIDGE DETECTED WARNING IS RECORDED ON (B)(4) 2012. THE PUMP IS ABLE TO POWER UP NORMALLY AND PASS "EZ-PRIME" STEPS SUCCESSFULLY. THE FORCE SENSOR CALIBRATION READING WAS TAKEN AND WAS FOUND ABOVE NORMAL SPECIFICATIONS. THE PUMP WAS OPENED, THE OLD TAPE AND THE FORCE SENSOR PINS WERE FOUND INTACT; NO DEFECT WAS FOUND TO THE FORCE SENSOR CIRCUIT. THE FORCE SENSOR RESISTANCE WAS TAKEN AND FOUND WITHIN SPECIFICATIONS. UNABLE TO DUPLICATE RECORDED "163" WARNING.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. THE PUMP REPORTEDLY DISPENSED INSULIN DURING THE LOAD CARTRIDGE STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |