FDA Adverse Event Other Summary report: N

INVACARE MEDICAL

MDR report key: 2867997 · Received October 26, 2012

Report

Report Number
3004766495-2012-00004
Event Type
Other
Date Received
October 26, 2012
Date of Event
February 8, 2012
Report Date
October 23, 2012
Manufacturer
DANYANG JUMAO HEALTHCARE EQUIPMENT CO.,LTD.
Product Code
IOR
Removal / Correction Number
KS1203001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPLACEMENT WHEEL SENT ON WARRANTY REPLACEMENT ORDER # (B)(4). REWORK ALL INVENTORY PRODUCTS, REPLACE THE CASTER BROKE. (THROUGH INVESTIGATION, NO FINISHED GOODS). OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. REWORK ALL SEMI-FINISHED PRODUCTS AT ASSEMBLY LINE. (THROUGH INVESTIGATION, NO SEMI-FINISHED PRODUCTS). OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. REMARK AND ISOLATE THE RAW MATERIAL (FRONT CASTER), THEN NOTICE IQC CHECK AGAIN. DO DROP TESTING FOR 4 PCS SAMPLE, IF PASSED, THE RAW MATERIAL (FRONT CASTER) CAN BE USED. OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. IPQC WILL INFORCE MONITORING THE PLASTIC INJECTION PARAMETERS. OPERATOR: (B)(4). FINISHED DATE: 06/25/2012.

Description of Event or Problem · 1

CUSTOMER ALLEGES ONE OF THE CASTER WHEELS HAD SPLIT COMPLETELY THROUGH. REORDER (B)(4). NO INJURY ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INVACARE MEDICAL MECHANICAL (MANUAL) WHEELCHAIR IOR DANYANG JUMAO HEALTHCARE EQUIPMENT CO.,LTD. V18RLR NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other