INVACARE MEDICAL
Report
- Report Number
- 3004766495-2012-00004
- Event Type
- Other
- Date Received
- October 26, 2012
- Date of Event
- February 8, 2012
- Report Date
- October 23, 2012
- Manufacturer
- DANYANG JUMAO HEALTHCARE EQUIPMENT CO.,LTD.
- Product Code
- IOR
- Removal / Correction Number
- KS1203001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
REPLACEMENT WHEEL SENT ON WARRANTY REPLACEMENT ORDER # (B)(4). REWORK ALL INVENTORY PRODUCTS, REPLACE THE CASTER BROKE. (THROUGH INVESTIGATION, NO FINISHED GOODS). OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. REWORK ALL SEMI-FINISHED PRODUCTS AT ASSEMBLY LINE. (THROUGH INVESTIGATION, NO SEMI-FINISHED PRODUCTS). OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. REMARK AND ISOLATE THE RAW MATERIAL (FRONT CASTER), THEN NOTICE IQC CHECK AGAIN. DO DROP TESTING FOR 4 PCS SAMPLE, IF PASSED, THE RAW MATERIAL (FRONT CASTER) CAN BE USED. OPERATOR: (B)(4). FINISHED DATE: 06/22/2012. IPQC WILL INFORCE MONITORING THE PLASTIC INJECTION PARAMETERS. OPERATOR: (B)(4). FINISHED DATE: 06/25/2012.
CUSTOMER ALLEGES ONE OF THE CASTER WHEELS HAD SPLIT COMPLETELY THROUGH. REORDER (B)(4). NO INJURY ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INVACARE MEDICAL | MECHANICAL (MANUAL) WHEELCHAIR | IOR | DANYANG JUMAO HEALTHCARE EQUIPMENT CO.,LTD. | V18RLR | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |