GENERATOR PS100-100 PULSAR
Report
- Report Number
- 1226420-2012-00089
- Event Type
- Malfunction
- Date Received
- December 11, 2012
- Date of Event
- November 9, 2012
- Report Date
- January 9, 2013
- Manufacturer
- MEDTRONIC ADVANCED ENERGY, LLC,
- Product Code
- MUL
- PMA / PMN Number
- K073057
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
(B)(4). EVAL, METHOD, RESULTS AND CONCLUSION: PRODUCT RECEIVED BY MANUFACTURER AND PENDING INSPECTION.
(B)(4). EVALUATION PROCESS: UNIT RECEIVED IN FAIR CONDITION AND INTERNAL VISUAL INSPECTION FOUND NOTHING MOVING OR BROKEN IN THE UNIT. UNIT DELIVERED RF ENERGY CORRECTLY INTO FIXED RESISTORS AND POWER WAS CHECKED IN ALL MODES AND POWER LEVELS AND ALL WERE WITHIN ACCEPTABLE LIMITS. UNIT WAS ALSO TESTED WITH PLASMABLADE INTO AN ANALOG SUBSTANCES AND EVENT AT THE HIGHEST POWER LEVELS NO HEATING OF THE PROBE WAS NOTED. (B)(4).
TIP OF PLASMABLADE CAUGHT FIRE. DOCTOR PERFORMED TWO T<(>&<)>A CASES AFTER THE INCIDENT WITH THE PEAK TNA PLASMABLADE WITHOUT INCIDENT.
TIP OF PLASMABLADE CAUGHT FIRE. DOCTOR PERFORMED TWO SUBSEQUENT TNA CASES AFTER THE INITIAL INCIDENT WITH THE PEAK TNA PLASMABLADE WITHOUT FURTHER INCIDENT.