FDA Adverse Event Malfunction Summary report: N

PLATO RTS V2.3

MDR report key: 286730 · Received July 13, 2000

Report

Report Number
1121753-2000-00001
Event Type
Malfunction
Date Received
July 13, 2000
Date of Event
May 23, 2000
Report Date
July 6, 2000
Manufacturer
NUCLETRON BV
Product Code
MUJ
Removal / Correction Number
NA
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

IN CERTAIN SITUATIONS WHEN ENTERING A BLOCK, THE PROGRAM INTERPRETS THE WHOLE FIELD AS BEING BLOCKED, AND A HIGH NUMBER OF MONITOR UNITS ARE CALCULATED. THE PROBLEM MAY OCCUR WHEN THE BLOCK CONTOUR CONTAINS A LARGE NUMBER OF POINTS AND/OR SOME OF THE POINTS ARE VERY CLOSE TOGETHER. IF THE PROBLEM OCCURS, A WARNING MESSAGE "HIGH ABSOLUTE M.U./MIN. AND/OR HIGH VALUES DETECTED" WILL BE DISPLAYED, AND CALCULATED MONITOR UNITS WILL BE EXCEPTIONALLY HIGH. IF THE WARNING IS IGNORED, UNPLANNED HIGH DOSES COULD BE DELIVERED DURING TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLATO RTS V2.3 RADIOTHERAPY TREATMENT PLANNING SYSTEM MUJ NUCLETRON BV RTS V2.3 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other