AMS GREENLIGHT HPS FIBER
Report
- Report Number
- 2937094-2012-01096
- Event Type
- Malfunction
- Date Received
- October 22, 2012
- Date of Event
- May 2, 2012
- Report Date
- June 1, 2012
- Manufacturer
- AMS
- Product Code
- GEX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE MFR ASSIGNED MODEL NUMBER 0010-2093 IS DESIGNATED FOR (B)(4) DISTRIBUTION. THIS REPORT IS BEING SUBMITTED AS THE MODEL 0010-2093 IS IDENTICAL IN DESIGN AND MANUFACTURE TO THE COMMERCIALLY AVAILABLE MODEL 0010-2090 ANGLED DELIVERY DEVICE, GREENLIGHT (K062719). FIBER ANALYSIS: THE FIBER CAP REMAINS INTACT AND ATTACHED AND DRILLED THROUGH; EXHIBITS DETRITUS, CHAR, DEVITRIFICATION, AND MELTED GLASS. CAP CONDITION WOULD RESULT IN REDUCED VAPORIZATION EFFICIENCY. THIS MAY CAUSE FORWARD FIRING. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT AT THE END OF THE PROCEDURE THE SIDE-FIRING FIBER WAS NOTICED TO HAVE COMMENCED FORWARD FIRING AND DECREASED VAPORIZATION AT 283, 750 JOULES DURING A PROSTATE PROCEDURE. AN ADD'L FIBER WAS NOT USED TO COMPLETE THE CASE. NO PT OR END USER INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMS GREENLIGHT HPS FIBER | POWERED LASER SURGICAL INSTRUMENT | GEX | AMS | 0010-2093 | 123J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GREENLIGHT HPS LASER SYSTEM| ACCESSORIES |