FDA Adverse Event Injury Summary report: N

FIRST DEVICE: 7.0MM REVERE DUAL OUTER DIAMETER

MDR report key: 2865134 · Received May 20, 2010

Report

Report Number
3004142400-2010-00014
Event Type
Injury
Date Received
May 20, 2010
Date of Event
April 21, 2010
Report Date
April 21, 2010
Manufacturer
GLOBUS MEDICAL, INC.
Product Code
NKB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A COMPREHENSIVE INVESTIGATION WAS IMMEDIATELY INITIATED ON RECEIPT OF THE COMPLAINT. A REVIEW WAS CONDUCTED OF ALL APPLICABLE MATERIAL RECORDS, MFG RECORDS, STORAGE RECORDS, AND DISTRIBUTION RECORDS. ALL RECORDS REVEALED THE PRODUCT WAS MANUFACTURED WITHIN SPECS, MAINTAINED, AND DISTRIBUTED IN ACCORDANCE WITH ALL FEDERAL, STATE AND OPERATING PROCEDURES. BASED ON RECORD REVIEW AND ALL AVAILABLE INFO, IT IS DETERMINED THE 7.0MM REVERE DUAL OUTER DIAMETER POLYAXIAL SCREW, 60MM DESIGN CONFORMED TO ALL APPLICABLE STANDARDS AND THERE WAS NO DEVIATION IN THE MANUFACTURE PROCESS. THERE IS NO INDICATION THAT THE ISSUE WAS A RESULT OF PRODUCT DEFECT OR MALFUNCTION. REF MFR REPORT: 3004142400-2010-00015 AND 00016.

Description of Event or Problem · 1

GLOBUS MEDICAL, INC RECEIVED NOTIFICATION FROM A SALES REP THAT GLOBUS MANUFACTURED SCREWS HAD BROKEN IN A PT AFTER IMPLANTATION AND WERE EXPLANTED. A MALE PT UNDERWENT ORIGINAL PLIF, L3-S1 SURGERY ON (B)(6) 2007 AND WAS IMPLANTED WITH A CONSTRUCT THAT INCLUDED REVERE DUAL OUTER DIAMETER POLYAXIAL SCREWS (DOD SCREWS) WITH CORRESPONDING RODS, ADJUSTABLE T-CONNECTOR, AND REVERE LOCKING CAPS. ON (B)(6) 2008, REVISION SURGERY OCCURRED ON THE PT BECAUSE THE RODS HAD SLIPPED OUT OF THE DOD SCREWS AND LOCKING CAPS AT S1. APPROX 26 MONTHS LATER, X-RAYS SHOWED THE PT HAD SOLID FUSION, BUT BOTH OF THE DOD SCREWS HAD BROKEN AT S1. REVISION SURGERY OCCURRED ON (B)(6) 2010 AND BOTH DOD SCREWS WERE REMOVED AND REVERE 8.5MM SCREWS WERE UTILIZED AT THE S1 LEVEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FIRST DEVICE: 7.0MM REVERE DUAL OUTER DIAMETER 7.0MM REVERE DUAL OUTER DIAMETER POLYAXIAL NKB GLOBUS MEDICAL, INC. 124.787 GMH230RA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SUSTAIN OBLIQUE SPACER| REVERE LOCKING CAPS| CP2 RODS| REVERE PEDICLE SCREWS| ADJUSTABLE T-CONNECTOR