CONCERTO CRT-D DR
Report
- Report Number
- 3004209178-2012-11417
- Event Type
- Injury
- Date Received
- December 10, 2012
- Report Date
- December 19, 2012
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. CONCOMITANT PRODUCTS: 4574 IMPLANTABLE PACING LEAD 2008 (B)(6); 4194 IMPLANTABLE PACING LEAD 2008 (B)(6). (B)(4).
PRODUCT EVENT SUMMARY : (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. DATA FROM THE DEVICE WAS ALSO COLLECTED AND ANALYZED. ANALYSIS OF THE DATA REVEALED THAT THERE WAS A LOW BATTERY VOLTAGE ALERT. TIME OF RECOMMENDED REPLACEMENT TIME (RRT) IN THE SAVE TO DISK WAS ON (B)(6) 2012. THE DEVICE RRT WAS LESS THAN OR EQUAL TO 2.62 VOLTS. THE WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BATTERY EQUALING 2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2012.
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO REACHING ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO REACHNG ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | C154DWK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD |