CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2012-17590
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- October 8, 2012
- Report Date
- October 8, 2012
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED AND THE PRIMARY ANALYSIS FINDING INDICATES THAT NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL END OF THE ELECTRODE WAS COVERED WITH BLOOD.
CORRECTED INFORMATION: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT IN DIFFERENT POSITIONS THE LEAD IMPEDANCE WAS HIGH. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI |