PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2012-05538
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 14, 2012
- Report Date
- November 16, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. CARTRIDGE PAN DISLODGED. THE ANALYSIS RESULTS THAT ONE DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND WITH A CARTRIDGE RELOAD WITHOUT PAN PRESENT. IN ADDITION A CARTRIDGE PAN WAS FOUND LODGE INSIDE THE CHANNEL. THE PAN WAS REMOVED FROM THE CHANNEL AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD. THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES NOTED. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. WHILE NO CONCLUSION COULD BE REACHED AS TO HOW THE PAN BECAME DISLODGE FROM THE RELOAD, IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A GASTRIC BYPASS PROCEDURE, AFTER THE FIRST FIRING THEY ATTEMPTED TO RELOAD THE DEVICE FOR A SECOND FIRING ON THE JEJUNUM. HOWEVER, THEY WERE UNABLE TO LOAD THE SECOND CARTRIDGE INTO THE JAWS. UPON INSPECTION, IT APPEARS THE METAL PAN WAS STUCK IN THE JAWS. THE DEVICE WAS REMOVED AND A NEW ONE WAS PULLED AND USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4CF9H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |