FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2864459
·
Received December 10, 2012
Report
- Report Number
- 3015876-2012-00884
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 13, 2012
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT FAILURE. PHYSIO REPLACED THE SYSTEM/MEMORY PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. THE REMOVED SYSTEM/MEMORY PCB ASSEMBLY WAS FURTHER EVALUATED IN THE FAILURE ANALYSIS CENTER. THE PCB ASSEMBLY PASSED FUNCTIONAL AND PERFORMANCE TESTING AND THERE WERE NO ABNORMALITIES FOUND. THE CONCLUSIVE CAUSE OF THE REPORTED FAILURE COULD NOT BE DETERMINED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD INTERMITTENTLY POWER ITSELF OFF. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |