FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2864459 · Received December 10, 2012

Report

Report Number
3015876-2012-00884
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 13, 2012
Report Date
November 13, 2012
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K102972
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT FAILURE. PHYSIO REPLACED THE SYSTEM/MEMORY PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. THE REMOVED SYSTEM/MEMORY PCB ASSEMBLY WAS FURTHER EVALUATED IN THE FAILURE ANALYSIS CENTER. THE PCB ASSEMBLY PASSED FUNCTIONAL AND PERFORMANCE TESTING AND THERE WERE NO ABNORMALITIES FOUND. THE CONCLUSIVE CAUSE OF THE REPORTED FAILURE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD INTERMITTENTLY POWER ITSELF OFF. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.