FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR® PLUS DEFIBRILLATOR

MDR report key: 2864283 · Received December 10, 2012

Report

Report Number
3015876-2012-00882
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
August 16, 2012
Report Date
September 10, 2018
Manufacturer
PHYSIO-CONTROL, INC. - 3015876
Product Code
MKJ
PMA / PMN Number
K033275
Removal / Correction Number
3015876-02/08/2013-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL FURTHER EVALUATED THE DEVICE AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO AN ELECTRONICALLY LEAKY FILTER, DESIGNATOR FL9, FROM THE ANALOG PCB ASSEMBLY WHICH CAUSED THE INTERNAL HLC BATTERY TO DEPLETE.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL HAS EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Additional Manufacturer Narrative · 1

THIS PRODUCT COMPLAINT REQUIRES A SUPPLEMENTAL MDR TO DOCUMENT THAT FIELD ACTION NUMBER 3015876-02/08/2013-001C IS RELEVANT TO THE REPORTED ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED TO A PHYSIO-CONTROL SERVICE REPRESENTATIVE THAT THE CUSTOMER'S DEVICE DID NOT POWER ON AND THAT THE CHARGE-PAK, ATTENTION AND SERVICE SYMBOLS WERE LIT. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Description of Event or Problem · 1

SUPPLEMENTAL MDR IS REQUIRED TO DOCUMENT THAT FIELD ACTION NUMBER 3015876-02/08/2013-001C IS RELEVANT TO THIS REPORTED ISSUE.