SWAN-GANZ THERMODILUTION PACEPORT CATHETER WITH AMC THROMBOSHIELD

MDR report key: 2864184 · Received December 10, 2012

Report

Report Number
2015691-2012-18808
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 12, 2012
Report Date
November 12, 2012
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
LDF
PMA / PMN Number
K803058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CATHETER IS EXPECTED TO BE RETURNED FOR EVALUATION; HOWEVER, IT HAS NOT YET BEEN RECEIVED.

Additional Manufacturer Narrative · 1

EVALUATION OF THE RETURNED PRODUCT REFUTED THE CUSTOMER REPORT OF "BALLOON IN THE TIP OF THE CATHETER WAS MISSING". BALLOON LEAKAGE WAS IDENTIFIED THROUGH A TEAR, ABOUT 0.040" IN LENGTH PROXIMAL OF THE DISTAL BOND. THE BALLOON WAS IMMERSED IN WATER AND PRESSURIZED WITH AIR TO VISUALLY DETERMINE THE SOURCE OF LEAKAGE. ALL THROUGH LUMENS WERE PATENT WITHOUT ANY LEAKAGE OR OCCLUSION. NO VISIBLE DAMAGE WAS OBSERVED ON CATHETER BODY. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED DURING THE EVALUATION. A REVIEW OF THE MANUFACTURING RECORDS INDICATED THAT THE PRODUCT MET SPECIFICATIONS UPON RELEASE. IT COULD NOT BE DETERMINED IF PROCEDURAL FACTORS OR DEVICE HANDLING MAY HAVE CONTRIBUTED TO THE EVENT REPORTED. NO ACTIONS WILL BE TAKEN AT THIS TIME.

Description of Event or Problem · 1

AS REPORTED, THE CUSTOMER INFORMED THAT THERE WAS NO BALLOON ON THE CATHETER. ADDITIONAL INFORMATION REVEALED THAT THE EVENT OCCURRED BEFORE USE ON PATIENT. DURING SET UP, THE STAFF NOTICED THAT THE BALLOON ON THE TIP OF THE CATHETER WAS MISSING. SOME "PARTS" THAT COULD HAVE BEEN A PART OF A BALLOON WERE OBSERVED. THE CATHETER WAS NEVER IN CONTACT WITH BLOOD OR THE PATIENT. THE CATHETER WAS HANDLED UNDER STERILE CONDITIONS SO THERE IS NO CONTAMINATION.