FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2864157 · Received December 10, 2012

Report

Report Number
3004209178-2012-91207
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 26, 2012
Report Date
November 26, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING REWIND DUE TO ENCODER OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO CONSTANT MOTOR ERROR ALARM.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED. THE BLOOD GLUCOSE READING WAS 153MG/DL. ASSISTED THE CUSTOMER TO PERFORM THE DISPLACEMENT TEST AND THE TEST FAILED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.