FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 2863866 · Received December 10, 2012

Report

Report Number
1826988-2012-00643
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 1, 2012
Report Date
November 18, 2012
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K111268
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 320MG/DL FROM HIS CONTOUR NEXT EZ METER, RE-TESTED ON HIS CONTOUR METER AND RECEIVED 87MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "D" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 7311 2EFEC02

Patients

Seq Age Sex Outcome Treatment
1