FDA Adverse Event
Malfunction
Summary report: N
CONTOUR NEXT
MDR report key: 2863866
·
Received December 10, 2012
Report
- Report Number
- 1826988-2012-00643
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 18, 2012
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K111268
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 320MG/DL FROM HIS CONTOUR NEXT EZ METER, RE-TESTED ON HIS CONTOUR METER AND RECEIVED 87MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "D" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7311 | 2EFEC02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |