STERRAD 50 STERILIZATION SYSTEM
Report
- Report Number
- 2084725-2012-00097
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 28, 2012
- Report Date
- November 28, 2012
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K981625
- Removal / Correction Number
- Z-0744-0746-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
BASED ON CAPA INVESTIGATION AND ASTM TESTING, OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD SYSTEMS. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS (FMEA), SYSTEM HAZARD AND USER MISUSE ANALYSIS (SHUMA), AND CORRECTIVE AND PREVENTATIVE ACTION (CAPA). THE DHR WAS REVIEWED AND INDICATED NO ANOMALIES WHICH COULD HAVE CONTRIBUTED THE CUSTOMER'S EXPERIENCED ISSUE. THE UNIT MET ALL SPECIFICATIONS AT THE TIME OF RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST SIX MONTHS (06/01/2012 TO 11/28/2012) DID NOT IDENTIFY ANY SIGNIFICANT TREND. THE TREND FOR THE PRODUCT MALFUNCTION CODE OF ODOR/SMELLS IDENTIFIED A BREACH DUE TO THE OUS REMEDIATION EFFORT CONTAINED IN A CAPA. A REVIEW OF THE COMPLAINT HISTORY DID NOT REVEAL A SIGNIFICANT TREND FOR US DATA OVER THE PAST TWELVE MONTHS (JANUARY 2013 THROUGH DECEMBER 2013). THE FMEA REVEALED THE RISK IS AT AN ACCEPTABLE LEVEL. THE SHUMA INDICATES THE RISK IS AS LOW AS REASONABLY PRACTICABLE FOR MILD (LIMITED) EXPOSURE TO ODOR OR ODORANTS. CAPA INVESTIGATION AS WELL AS THE ASTM TESTING PERFORMED INDICATED OIL DEGRADATION CAN CAUSE ODOR/SMELLS FOR STERRAD® SYSTEMS. THE OIL FILTER HOUSING WAS RETURNED AND TESTED. THE OIL FILTER HOUSING WAS MISSING THE "HOUSING WINDOW," ALLOWING MIST/ODOR TO ESCAPE. THE REASON FOR RETURN OF THE OIL FILTER HOUSING WAS CONFIRMED. THE REPORTED ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. THE ISSUE WILL BE TRACKED AND TRENDED. NO FURTHER INVESTIGATION IS NECESSARY AT THIS TIME.
FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. A PM2 WAS PERFORMED AND THE OIL FILTER HOUSING, COVER INSULATION AND FILTER WERE REPLACED.
A CUSTOMER REPORTED AN EVENT OF A CHEMICAL ODOR EMITTING FROM THE STERRAD 50. THERE WAS NO REPORT OF HUMAN REACTION. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2012.