FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2863372 · Received December 10, 2012

Report

Report Number
1416980-2012-06784
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 15, 2012
Report Date
November 15, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. PER THE CUSTOMER, THIS DEVICE IS NOT AVAILABLE FOR EVALUATION. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED, AND THE ROOT CAUSE COULD NOT BE DETERMINED. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE LEAKED AFTER FILLING. THE DEVICE WAS FILLED WITH A 250-ML SOLUTION OF 90 MG PAMIDRONATE IN NORMAL SALINE. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 12F077

Patients

Seq Age Sex Outcome Treatment
1