FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY SPATULA

MDR report key: 2863358 · Received December 10, 2012

Report

Report Number
2955842-2012-01204
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 12, 2012
Report Date
November 12, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. THE COMPLAINT OF THE INSULATION BEING CRACKED WAS CONFIRMED. THE CERAMIC SLEEVE HAS A SECTION BROKEN OFF AT THE DISTAL END. PIECES OF THE CERAMIC ARE MISSING FROM THE SPATULA, EXPOSING THE YAW PULLEY AND SPATULA SHANK. THE SLEEVE ALSO EXHIBITS CRACKING BELOW THE BROKEN SECTION. EVIDENCE WAS NOT CONCLUSIVE, BUT DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. THERE WERE 8 USES LEFT. NO OTHER DAMAGE WAS FOUND. INTUITIVE SURGICAL CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS OF THE DATE OF THIS REPORT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVE. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE INSULATION OF THE PERMANENT CAUTERY HOOK INSTRUMENT WAS CRACKED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERMANENT CAUTERY SPATULA ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420184-06 M10120713 133

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES