FDA Adverse Event Malfunction Summary report: N

SMALL GRASPING RETRACTOR INSTRUMENT

MDR report key: 2863273 · Received December 10, 2012

Report

Report Number
2955842-2012-01215
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 7, 2012
Report Date
November 7, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT OF INSTRUMENT WIRES ARE SNAPPED AT THE DISTAL END WAS CONFIRMED. THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL REMAINED IN THE CLEVIS. CLEVIS DOES NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, DURING SET UP, THE SURGICAL STAFF REPORTED WIRES SNAPPED ON THE SMALL GRASPING RETRACTOR INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SMALL GRASPING RETRACTOR INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420318-02 M10120724 757

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES