PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2863225
·
Received December 10, 2012
Report
- Report Number
- 3004209178-2012-91136
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- October 22, 2012
- Report Date
- November 28, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. THE MOTOR WAS TESTED AND PASSED THE TEST. THE DEVICE HAD MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR SEVERAL TIMES DURING BASAL DELIVERY. THE BLOOD GLUCOSE READING WAS 156MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE MOTHER TO RUN THE DISPLACEMENT AND SELF TEST AND THEY PASSED. NO FURTHER INFORMATION WAS PROVIDED.