FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2863205
·
Received December 10, 2012
Report
- Report Number
- 3004209178-2012-91170
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- September 23, 2012
- Report Date
- October 26, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY AND CONSTANT TONE AS A RESULT OF A FAULTY COMPONENT ON THE INTERFACE BOARD. UNABLE TO CONFIRM THE ALARM OR THE BLANK DISPLAY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED, AND SHE WAS NOT ABLE TO CLEAR THE ALARM. THE CALLER REMOVED THE BATTERY FOR FIVE MINUTES, BUT THE DISPLAY DID NOT RETURN. INSTRUCTED THE CUSTOMER TO REMOVE THE BATTERY FOR TWO HOURS, BUT THE DEVICE TURNED OFF AGAIN SEVERAL TIMES AFTER THE INSULIN PUMP WAS RESET. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |