FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2863202 · Received December 10, 2012

Report

Report Number
3004209178-2012-91133
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
September 20, 2012
Report Date
November 29, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST AND REWINDS PROPERLY. HOWEVER, THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER COULD NOT REWIND THE INSULIN PUMP. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWL

Patients

Seq Age Sex Outcome Treatment
1