FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2862842 · Received December 10, 2012

Report

Report Number
1416980-2012-06669
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 1, 2012
Report Date
November 15, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED PROBLEM OF BATTERY LOW ALARM WAS CONFIRMED DURING DEVICE EVALUATION. SERVICE DETERMINED THAT THE CAUSE WAS DUE TO DEPLETED/DEFECTIVE MAIN BATTERIES. THE MAIN BATTERIES WERE REPLACED TO RESOLVE THE ISSUE. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BAXTER (B)(4) OF A FLOGARD INFUSION PUMP WITH A BATTERY LOW ALARM. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION REPORTED IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1