INFUSOR
Report
- Report Number
- 1416980-2012-06740
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 13, 2012
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K011317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THAT THE COIL CAP SEPARATED FROM THE COVER. THE ROOT CAUSE WAS DETERMINED TO BE A MANUFACTURING DEFECT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
BAXTER (B)(4) RECEIVED A REPORT THAT A MULTIRATE INFUSOR HAD A COIL CAP POP OFF DURING FILLING. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 11C089 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |