FDA Adverse Event Summary report: N

POWERED WHEELCHAIR

MDR report key: 2862763 · Received December 10, 2012

Report

Report Number
1531186-2012-01677
Date Received
December 10, 2012
Report Date
December 7, 2012
Manufacturer
ANI [ A&I ] GROUP TAIWAN
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

CALLED (B)(6), IN WHICH SHE STATED THAT SHE RECEIVED A CALL ON (B)(6) 2012 FROM END USER. (B)(6) ALLEGES THAT THE END USER STATED HE WAS IN HIS CHAIR, WENT DOWN THE RAMP AT HIS RESIDENCE AND FELL OVER WITH THE CHAIR. SHE STATED THE END USER DID NOT HAVE ANY SERIOUS INJURIES, ONLY SCRATCHES. (B)(6) ALLEGES THE END USER DID NOT SEEK MEDICAL ATTENTION. SHE SENT TECHNICIAN TO PATIENTS RESIDENCE IN WHICH TECHNICIAN ALLEGES THAT END USERS' RAMP IS TOO STEEP AND BECAUSE OF THE MOMENTUM OF GOING DOWN RAMP THIS CAUSED THE CASTORS TO ROTATE. (B)(6) STATED THE TECHNICIAN ADVISED END USER TO REPLACE RAMP. (B)(4).

Description of Event or Problem · 1

PROVIDER STATES FRONT CASTER ROTATED IN THE FRAME MOUNT WHILE USER WAS DRIVING CHAIR, CAUSING HIM TO FALL FROM CHAIR. REPORTER ALLEGES END USER WAS IN HIS CHAIR WENT DOWN THE RAMP AT HIS HOME AND FELL WITH THE CHAIR. NO INJURY, BUT SCRATCHES, AND DID NOT SEEK MEDICAL ATTENTION. THE TECHNICIAN ALLEGES THAT END USER'S RAMP IS TOO STEEP AND BECAUSE OF THE MOMENTUM OF GOING DOWN RAMP, THIS CAUSED THE CASTORS TO ROTATE. (B)(6) STATED THE TECHNICIAN ADVISED END USER TO REPLACE RAMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI ANI [ A&I ] GROUP TAIWAN IN88AHANFR

Patients

Seq Age Sex Outcome Treatment
1 53 YR Other