INTERMATE
Report
- Report Number
- 1416980-2012-06737
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 13, 2012
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- PMA / PMN Number
- K910425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A RUPTURED RESERVOIR WAS CONFIRMED. VISUAL EXAMINATION OF THE UNIT DETERMINED THAT THE BLADDER RUPTURED IN A NON-FOOTED POSITION. THE ROOT CAUSE COULD NOT BE DETERMINED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. THE INCORRECT MANUFACTURING DATE WAS SUBMITTED IN THE INITIAL MEDWATCH.
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.
BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE HAD A RESERVOIR RUPTURE BEFORE PATIENT USE. THE DEVICE WAS BEING FILLED WITH A 100 ML SOLUTION OF GLUCOSE. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTERMATE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - IRVINE | 12C079 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |