FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 2862587 · Received December 10, 2012

Report

Report Number
2024168-2012-07761
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 13, 2012
Report Date
November 14, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE SHAFT SEPARATION WAS CONFIRMED. THE FAILURE TO ADVANCE/CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE MID CIRCUMFLEX ARTERY WITH MILD TORTUOSITY AND NO CALCIFICATION. AFTER PRE-DILATATION WAS PERFORMED, THE 2.75 X 15 MM XIENCE PRIME STENT DELIVERY SYSTEM (SDS) WAS ADVANCED, BUT THE SDS SHAFT SEPARATED 8 CM FROM THE PROXIMAL HUB BEFORE THE SDS COULD CROSS TO THE LESION. IT WAS NOTED THAT UNDUE FORCE WAS NOT APPLIED. THE DISTAL SECTION OF THE SDS WAS EASILY REMOVED WITHOUT INCIDENT. THE STENT WAS NOT DEPLOYED. ANOTHER STENT WAS THEN DEPLOYED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2030241

Patients

Seq Age Sex Outcome Treatment
1 66 YR